Direct Reflection
The sensor must be mounted according to the installation drawing
so that the reflected beam hits the receiver.
Mount the sensor using the through-bores and three M4 screws
from the accessories.
MR SMR + 0.5 MR alpha
2 (0.08) 25 (0.98) 20.5 °
5 (0.20) 26.5 (1.04) 20 °
10 (0.39) 35 (1.38) 17.5 °
20 (0.79) 50 (1.97) 13.8 °
SMR = Start of measuring range | MR = Measuring range
Dimensions in mm (inches)
Mounting steps
- Switch on the power supply on the sen-
sor.
- Observe how the color of the Status
LED on the sensor changes.
- Position a reflective measuring object
within the measuring range..
Measuring
- Push the optional mounting aid between
the sensor and the measuring object.
- The Status LED lights up yellow.
- Fasten the sensor using three M4 screws.
Remove the mounting aid between the sen-
sor and the measuring object.
33.5
4
4.54
10
alpha
48
2
MR (dir.R)
Through-holes
3x ø4.5 (dia. 0.18)
for M4 screws
Limits for free
Space
13.5
67 (2.6)
89 (3.5)
97 (3.8)
SMR + 0.5 MR
Fig. 9 Dimensional drawing, free space for diffuse reflection