Page 29
Installation | Housing S
optoNCDT 2300
Diffuse Reflection
If the laser beam does not strike the object surface at a
perpendicular angle, measurements might be inaccu-
rate.
Mount the sensor using the through-bores and
three M4 screws from the accessories.
MR SMR Y
2 (.08) 24 (.94) 1.5 (.06)
5 (.20) 24 (.94) 3.5 (.14)
10 (.39) 30 (1.18) 6.5 (.26)
20 (.79) 40 (1.57) 10.0 (.39)
50 (1.97) 45 (1.77) 23.0 (.91)
100 (3.94) 70 (2.76) 33.5 (1.32)
MR = Measuring range
SMR = Start of measuring range
Dimensions in mm (inches)
67 (2.64)
33.5 (1.32)
4
4.5
89 (3.50)
SMR
75 (2.95)
4
Y
10
MR
48
2
Keep this area free from
other light sources and/or
their reflections
Through-holes
ø4.5
for M4 screws
15 (0.59)
7 7
Limits for free
space
Laser
13.5
Fig. 8 Dimensional drawing, free space for diffuse reflection