Control Menu, Set Sensor Parameter
optoNCDT 2300
7.6.1.3 Signal Processing -
Trigger for Acquiring Values
The current array signal is pro-
cessed only after a valid trigger
event, and it is used to calculate
the measurement values. The
measurement values are then
forwarded for further calculation
(e.g. averaging or statistics) and
for output via a digital interface.
When calculating averages or
statistics, measurement values
recorded immediately before the
trigger event cannot be included;
instead older measurement values
are used, which were recorded
during previous trigger events.
A
A
Video signal
Distance calculation
Error identification
Video averaging
Masking
Refractive index correction
Error logging
Found error?
no
yes
Spike correction
Difference / Thickness
Measurement averaging
Statistics
Hold last value
Relative measurements after
Zeroing / Mastering
Trigger signal?
no
yes
Data reduction
GetValue
Data output
No data output
Trigger
Trigger processing