Appendix| ASCII Communication with Sensor
optoNCDT 2300
A 6.5.2 Select Measurement Values to be Output
Setting the values to be output via the RS422 and Ethernet interface. The maximum output rate via the Ethernet interface depends on
the number of output values .
A 6.5.2.1 Request Data Selection
GETOUTINFO_ETH
GETOUTINFO_RS422
The commands list all selected output data for the interfaces Ethernet or RS422. The sequence shown corresponds to the output
sequence.
A 6.5.2.2 Data Selection Displacement Measurement
OUTDIST_RS422 NONE|([DIST1][DIST2])
Setting, which displacement values are output through RS422.
- NONE: No output of a displacement
- DIST1: Output of displacement 1
- DIST2: Output of displacement 2 (only possible, if thickness measurement is selected, see Chap. A 6.4.1.1)
It can also be output two displacements. If displacement measurement is selected for “MEASMODE”, the selected peak (see “MEA-
SPEAK”) is output as displacement value for “DIST 1”.
If thickness measurement is selected for “MEASMODE”, “DIST 1” represents the displacement value for the 1st peak.
A 6.5.2.3 Data Selection Thickness Measurement
OUTTHICK_RS422 NONE|[THICK12]
Defines, which calculated layer thickness is output via RS422.
- NONE: No output of calculated layer thickness
- THICK12: Output of the layer thickness between displacement 1 and 2.
This command is available only in the MEASMODE THICKNESS setting.