Page 139
Appendix| ASCII Communication with Sensor
optoNCDT 2300
A 6.4.3.4 Edit Material Table
MATERIALEDIT <Name> <Description> (nF))
Add or edit material.
- Name: Name of material
- Description: Description of material
- nF: refraction index nF at 670 nm (min: 1.0, max: 4)
The material table can contain a maximum of 20 materials.
A 6.4.3.5 Delete Material Table
MATERIALDELETE <Name>
Deletes material from material table.
A 6.4.4 Measurement Value Processing
A 6.4.4.1 Averaging of Measurement Value
AVERAGE NONE|MOVING|RECURSIVE|MEDIAN [<Averaging depth>]
The averaging value always affects all to be output displacement and difference values.
- MOVING: Moving averaging value (averaging depth 2, 4, 8, 16, 32, 64 and 128 possible)
- RECURSIVE: Recursive averaging value (averaging depth 1 up to 32768 possible)
- MEDIAN: Median (averaging depth 3, 5, 7 and 9 possible)
A 6.4.4.2 Spike Correction
SPIKECORR [ON|OFF[[<Number of evaluated measured values>][[<Tolerance range in mm>][<Number of cor-
rected values>]]]
Spike correction is not enabled in the factory default settings.
Factory settings Min Max
Number of measured values evaluated 3 1 10
Tolerance range in mm 0.1000000 0.0000000 100.0000000
Number of corrected values 1 1 100