Control Menu, Set Sensor Parameter
optoNCDT 2300
7.6.1.4 Signal Processing -
Trigger for Outputting all Values
This setting acts the same as any
selected trigger type; it supports
the same trigger behavior settings
as any other trigger type. The dif-
ference is that the trigger settings
do not affect signal processing. In-
stead, trigger data are transferred
using Bit 15 of the status word.
Video signal
Distance calculation
Error identification
Video averaging
Masking
Refractive index correction
Error logging
Found error?
no
yes
Spike correction
Difference / Thickness
Measurement averaging
Statistics
Hold last value
Relative measurements after
Zeroing / Mastering
Input trigger?
yes
Data reduction
TriggerState
GetValue
Data output
Trigger
Trigger processing