Page 118
Appendix| PC2300-0.5/Y
optoNCDT 2300
A 3 PC2300-0.5/Y
The PC2300-0.5/Y cable splits the sensor signals to an RJ45 female connector (Ethernet) and a cable with open leads. Cable length is
0.5 m.
Signal 15-pin Sub-D connector Open leads RJ45 connector
+
U
B
1 white
GND 9 brown
+Laser on/off
2
1
green
- Laser on/off
10
1
yellow
Sync-in/out 3 grey
/Sync-in/out 11 pink
RxD-RS422 4 blue
/RxD-RS422 12 red
TxD-RS422 5 black
/TxD-RS422 13 violet
Shield Housing Cable screen
Tx - Ethernet 6 1
/Tx - Ethernet 14 2
Rx - Ethernet 7 3
/Rx - Ethernet 15 6
Shield Housing Housing
Cable shield is provided with a ferrule. The strands of RS422 and synchronization are cut blunt.
1) +U
B
and +Laser on/off are connected together. GND and –Laser on/off are connected together.