EasyManua.ls Logo

MICRO-EPSILON optoNCDT 2300 - A 6.3.3 Triggering; A 6.3.3.1 Trigger Selection; A 6.3.3.2 Effect of the Trigger Input; A 6.3.3.3 Trigger Level

MICRO-EPSILON optoNCDT 2300
206 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 132
Appendix| ASCII Communication with Sensor
optoNCDT 2300
A 6.3.3 Triggering
Trigger-input serves also as synchronous input, which means level and edge triggering is only alternatively possible to sync mode,
see Chap. 7.6.3 (Synchronization).
A 6.3.3.1 Trigger Selection
TRIGGER NONE|EDGE|PULSE|SOFTWARE TERMON|TERMOFF
- NONE: No triggering
- PULSE: Level triggering
- EDGE: Edge triggering
- SOFTWARE: Software triggering
Set timing for sync- / trigger-input:
- TERMON: Timing (type 120 OHM)
- TERMOFF: No timing
The TRIGGER command always expects two parameters. If software triggerung is used, the second parameter is not evaluated.
A 6.3.3.2 Effect of the Trigger Input
TRIGGERAT [INPUT|OUTPUT]
- INPUT: Triggers the measurement value recording. Measurement values immediately before the trigger event are not included
when calculating the mean value. Instead, older values are used which were output during previous trigger events.
- OUTPUT: Triggers the measurement value output. Measurement values immediately before the trigger event are included when
calculating the mean.
A 6.3.3.3 Trigger Level
TRIGGERLEVEL HIGH|LOW
- HIGH: Edge triggering: Rising edge, level triggering: High-active
- LOW: Edge triggering: Falling edge, level triggering: Low-active

Table of Contents