Page 197
Appendix| Control Menu
optoNCDT 2300
Measuring rate 1.5 / 2.5 / 5 / 10 / 20 / 30 / 49,140 kHz
For dark or bright measuring objects a slower
measuring rate may be required. The con-
trol may however expose no longer than the
measuring rate allows. The measurement
range of the sensor is reduced at 49,140 kHz.
Averaging,
error handling,
statistics
Video averaging
No averaging /
Recursive 2 / 4 / 8
Moving 2 / 3 / 4
Median 3
Video averaging is effected before the cal-
culation of the displacement or thickness.
Recommended for very small peaks respec-
tively to receive more valid data.
Measurement
averaging
No averaging
Moving N values 2 / 4 / 8 ... 128 Value
Indication of averaging mode. The averaging
number N indicates the number of consecu-
tive measurement values to be averaged in
the sensor.
Recursive N
values
2 ... 32768 Value
Median N values 3 / 5 / 7 / 9 Value
Error handling
Error output, no measurement value Sensor emits error value.
Keep last value 0 ... 1024 Value
If no valid measurement value is determined,
the last valid value can be hold for a certain
period, that is, output repeatedly. The last
valid value is kept indefinitely at „0“.
Keep last value infinitely The last valid value is output indefinitely.
Spike correction
No
Yes
Evaluation length,
1 ... 10
Value
This filter removes individual very high spikes
from a relatively constant course of measure-
ment value. Smaller spikes are preserved.
Max. tolerance
range [mm], 0 ...
100
Value
Number of correct
value, 1 ... 100
Value