10-11
3. IBP: If the ECG signal is seriously interfered, you can select IBP which means PR will be
calculated by IBP module. In this case, PULSE is displayed to the right of the ECG label
while the PR reading is displayed below. The monitor activates pulse beeps from IBP
module.
4. AUTO: The monitor determines the heart rate source depending on the signal quality. The
priorities for heart rate calculation are: ECG, SpO
2
, IBP.
5. ALL: The monitor displays both the HR and the PR values. PR value is displayed at the
right of SpO
2
label or IBP label, SpO2 takes priority of IBP. The monitor will alarm for both
abnormal HR and PR. HR is given priority in determining the source of the beep tone. If HR
is not available, the sound will be from the PR.
HR CHANNEL CH1: The heart rate is determined by the first ECG
waveform;
CH2: The heart rate is determined by the second ECG
waveform;
AUTO: The monitor automatically selects an optimal
channel to calculate the heart rate.
This option is available in Mindray DS algorithm package only. For Mortara algorithm package,
the monitor uses both channels to calculate the heart rate simultaneously.
LEAD TYPE Options: 3 LEADS, 5 LEADS;
SWEEP Options: 12.5, 25.0 and 50.0 mm/s;
ECG DISPLAY
When LEAD TYPE is set to 5 LEADS, you can select one of the following options:
1. NORMAL DISPLAY: The monitor displays two ECG waveforms.
2. MULTI-LEADS DISPLAY: The monitor displays six ECG waveforms, which occupies the
whole waveforms area.
3. HALF-SCN MULTI-LEADS: The monitor displays six ECG waveforms, which only
occupies four waveform positions.
When LEAD TYPE is set to 3 LEADS, only NORMAL DISPLAY can be selected for the ECG
DISPLAY.
Other Setup
Select OTHER SETUP >> in the ECG SETUP menu. The following menu appears. The options
available in the menu vary with each algorithm package.

Table of Contents

Other manuals for Mindray DPM 4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DPM 4 and is the answer not in the manual?

Mindray DPM 4 Specifications

General IconGeneral
BrandMindray
ModelDPM 4
CategoryMedical Equipment
LanguageEnglish

Related product manuals