10-12
Figure 10-9 Other ECG Setup (Mindray DS Algorithm Package)
Figure 10-10 Other ECG Setup (Mortara Algorithm Package)
In this menu, you can select any of the displayed option. All options are listed below. Your
algorithm package may not include some of these options.
SMART LEAD
OFF
ON: If there is a LEAD OFF in the HR-derived channel, the
system will automatically switch to another lead to restore the
display of the ECG waveform in the HR-derived channel.
OFF: The SMART LEAD OFF is turned off.
When ECG is in NORMAL DISPLAY mode, if the SMART LEAD OFF is configured ON and
there is a LEAD OFF in the HR-derived channel, the system will automatically switch to another
lead to restore the display of the ECG waveform in the HR-derived channel. When the LEAD
OFF condition is corrected, the leads are automatically switched back. However, if you manually
switch the HR-derived channel, change the ECG display mode, or turn off SMART LEAD OFF
before the LEAD OFF condition is corrected, the leads will not be switched back.