12-7
îš„ SYS ALM HI Determines the upper limit of the systolic pressure.
îš„ SYS ALM LO Determines the lower limit of the systolic pressure.
îš„ MEAN ALM HI Determines the upper limit of the mean pressure.
îš„ MEAN ALM LO Determines the lower limit of the mean pressure.
îš„ DIA ALM HI Determines the upper limit of the diastolic pressure.
îš„ DIA ALM LO Determines the lower limit of the diastolic pressure.
If a measured pressure crosses a preset upper or lower alarm limit, an alarm will be triggered.
The NIBP alarm limits are as follows:
Adult Pediatric Neonate
Patient type
Upper
Limit
Lower
Limit
Upper
Limit
Lower
Limit
Upper
Limit
Lower
Limit
Systolic pressure
(mmHg)
(lower
limit+2)
~270
40~(upper
limit-2)
(lower
limit+2)
~200
40~(upper
limit-2)
(lower
limit+2)
~135
40~(upper
limit-2)
Mean pressure
(mmHg)
(lower
limit+2)
~230
20~(upper
limit-2)
(lower
limit+2)
~165
20~(upper
limit-2)
(lower
limit+2)
~110
20~(upper
limit-2)
Diastolic pressure
(mmHg)
(lower
limit+2)
~210
10~(upper
limit-2)
(lower
limit+2)
~150
10~(upper
limit-2)
(lower
limit+2)
~100
10~(upper
limit-2)
The alarm limits in factory configuration (Unit: mmHg):
Patient type Adult Pediatric Neonate
Sys alarm high 160 120 90
Sys alarm low 90 70 40
Mean alarm high 110 90 70
Mean alarm low 60 50 25
Dia alarm high 90 70 60
Dia alarm low 50 40 20

Table of Contents

Other manuals for Mindray DPM 4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DPM 4 and is the answer not in the manual?

Mindray DPM 4 Specifications

General IconGeneral
BrandMindray
ModelDPM 4
CategoryMedical Equipment
LanguageEnglish

Related product manuals