12-8
DISPLAY WAY 1 GROUP: The NIBP parameter area only displays a group
of NIBP values obtained in the latest measurement.
GROUPS: Multiple groups of measurement values are
displayed at the lower left corner of the screen as shown
below.
NS NM ND PR DATE TIME
108 84 70 68 2006-6-12 16:06:00
108 80 70 67 2006-6-12 16:11:00
108 80 70 72 2006-6-12 16:16:00
UNIT Options: mmHg, kPa;
INTERVAL Select MANUAL to set the monitor to manual NIBP
measurement mode, or select from the time options to
determine the interval between automatic measurements.
Optional intervals: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180,
240, and 480MIN.
RESET Select this option to restore the initial settings of the
pressure pump. If the monitor fails to give a visual
indication when the pressure pump is working improperly,
selecting this option activates a self-test procedure, and
restores the monitor to normal performance.
CONTINUAL Select this option to start a continuous measurement for five
minutes.
CONFIG >> You can select this option to access the NIBP CONFIG
menu, in which you may select FACTORY CONFIG or
USER CONFIG. After finishing your selection and exiting
the menu, the system pops up a dialog box asking for your
confirmation.