B-1
B EMC
The equipment meets the requirements of EN 60601-1-2.
NOTE
z Use of accessories, transducers, and cables other than those specified may result
in increased emission and/or decreased immunity of the equipment.
z The equipment should not be used adjacent to or stacked with other equipment,
and if adjacent or stacked use is necessary, the equipment should be observed to
verify normal operation in the configuration in which it will be used.
z The equipment needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided
below.
z The equipment may be interfered with by other equipment, even if that other
equipment complies with CISPR emission requirements.
z Operation of the device, in the case that the patient physiological signal is lower
than the minimum amplitude and/or value specified in the product
specifications, may cause inaccurate results.
TABLE 1
Guidance and declaration
— electromagnetic emissions
The equipment is intended for use in the electromagnetic environment specified below.
The customer or the user of the equipment should assure that it is used in such an environment.
Emissions test Compliance
Electromagnetic environment
— guidance
RF emissions
CISPR 11
Group1 The equipment uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class A
Harmonic
Emissions
IEC61000-3-2
Class A
Voltage
Fluctuations/Flicker
Emissions IEC
61000-3-3
Compliance
P
st
,T
dt
(ms)
D
max
(%)
Dc (%)
The equipment is suitable for use in all establishments
other than domestic and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.