A long instrument flushing cycle using the water from the dental unit
waterlines is automatically started after opening the tap. The long
flushing takes 4 minutes by default in a dental unit with cup fill and bowl
rinse functions.
To extend the duration, contact your Planmeca dealer.
The picture below shows the water tap when it is open.
7. After the long flushing cycle, make sure that the water coming from the
instruments and cup fill line is clear.
If the water is blue, run water through all water consuming instruments
and the cup fill line until the water is clear, and increase the flushing
time.
NOTE
To extend the duration, contact your Planmeca dealer.
After this, the unit is ready for normal use.