Help messages
HELP
CODE
MEANING ACTION OR EXPLANATION
H 23 Automatic/manual chip blow cannot be used
when sterile water function is selected.
H 24 Programming of selected setting is disabled. The programming of the selected setting is
disabled.
Contact your Planmeca dealer.
H 25 Planmeca Compact e help message:
Patient chair cannot be driven into
automatic chair position.
Contact your Planmeca dealer.
H 26 Drive chair upwards, then lower backrest. If
patient chair is attached to unit, chair's
automatic positions are not switched on.
Drive the chair upwards before lowering the
backrest. If this help message is shown
while the patient chair is attached to the
unit, it means that you have forgotten to
switch on the chair’s automatic positions. To
clear the message, put the operation mode
switch on the cuspidor side of the chair base
in the “AUTO” position (switch up).
H 27 Drive backrest upwards, then lower chair. If
patient chair is attached to unit, chair's
automatic positions are not switched on.
Drive the backrest upwards before lowering
the chair. If this help message is shown
while the patient chair is attached to the
unit, it means that you have forgotten to
switch on the chair’s automatic positions. To
clear the message, put the operation mode
switch on the cuspidor side of the chair base
in the “AUTO” position (switch up).
H 28 Planmeca Compact WE's help message. Connect the lifter cable.
H 29 Drive chair downwards, then lower backrest.
If patient chair is attached to unit, chair's
automatic positions are not switched on.
Drive the chair downwards before lowering
the backrest. If this help message is shown
while the patient chair is attached to the
unit, it means that you have forgotten to
switch on the chair’s automatic positions. To
clear the message, put the operation mode
switch on the cuspidor side of the chair base
in the “AUTO” position (switch up).
H 30 Close water tap. Fill container with
disinfectant up to groove and attach to unit.
H 31 Switch unit off, leave disinfectant in unit
overnight (min. 8 h).
When unit is turned back on, disinfection
procedure continues.
H31.1 Remove container, fill it up to groove with
water and attach to unit.
H31.2 Attach container filled with water.
H 32 Container flushing completed. Remove container from unit.
H 33 Open water tap.
H 34 No instruments selected when starting
instrument flushing or waterline cleaning.
Select at least one instrument and syringe
(remove from holder, bend the instrument
arms).
31 Help and error messages
258 Planmeca Compact i User's manual

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca Compact i Touch v2 and is the answer not in the manual?

Planmeca Compact i Touch v2 Specifications

General IconGeneral
BrandPlanmeca
ModelCompact i Touch v2
CategoryMedical Equipment
LanguageEnglish

Related product manuals