Help messages
HELP
CODE
MEANING ACTION OR EXPLANATION
H 35 Instruments are locked for safety reasons. Select instruments, place them in flushing
holder, and press Instrument flushing button
to start instrument flushing.
H35.1 Waterline cleaning cancelled or failed.
Remove container.
Make sure water tap is open. Select
instruments, place them in flushing holder
and press Instrument flushing button to start
instrument flushing.
H 36 Flushing ready. Return instruments to
instrument console.
Also bend and release syringe's instrument
arm once.
H36.1 Instrument flushing completed, but water
flow was not detected for one or more
instruments.
Run instruments for a while with spray water
to ensure instrument flushing is sufficient.
H36.8 Low water pressure during final flushing. Perform long instrument flushing to ensure
instruments are flushed.
H 37 Waterline cleaning has been interrupted.
H 38 Low water pressure during final rinsing. When waterline cleaning is finished, perform
long instrument flushing to ensure
instruments are flushed.
H 39 Syringe misplaced or its flow not detected in
waterline cleaning.
Return syringe to console. Then, select
syringe (bend instrument arm) again and
place it carefully in flushing holder.
H 40 Too many Bien-Air MX brushless
micromotors.
A maximum of two is allowed.
H 41 Insert suction tubes to suction tube cleaning
holder, add disinfectant and press green
start button (upper button).
H 41.1 Insert suction tubes to suction tube cleaning
holder.
H 41.2 Remove suction tubes from holder to start
suction cleaning.
H 42 Suction cleaning completed. Return suction
tubes to suction holder and close suction
cleaning cover.
H 42.1 Suction cleaning completed. Return suction
tubes to suction holder.
H 43 Suction cleaning cancelled. Return suction
tubes to suction holder.
H 44 Incorrect suction tube amount detected.
Return them to suction holder, place them
back to suction tube cleaning holder and
start cleaning.
H 45 Set WMS to Domestic Water Supply -mode
and wait until WMS bottle is full. Then start
suction cleaning.
Suction cleaning can not be used when
WMS is set to Bottled Water -mode.
H 46 Suction use is forbidden or there are no
suction tubes in dental unit configuration.
The value of service mode n.83 is 0.
31 Help and error messages
User's manual Planmeca Compact i 259