Help messages
HELP
CODE
MEANING ACTION OR EXPLANATION
H 47 Suction disinfectant pressure is low. (In
automatised suction cleaning: Fill suction
disinfectant container).
H 47.1 Suction disinfectant level low. Fill suction
disinfectant container.
H 47.2 Suction disinfectant container filled.
H 48 WMS must be set to Bottled water off and
Bottle on -modes before starting suction
cleaning.
Bottled water off = domestic water is used
Bottle on = bottle is used; domestic water is
run through the bottle
H 50 No cup in cup holder. Sensor does not
recognise cup when pressing cup fill button.
Place a cup on the cup holder. Or change
the setting of service mode n 115.
H 51 Bowl is above patient chair when driving
chair up.
If the unit has a bowl, move the bowl to the
rest position. If the unit has an OP delivery
arm safety switch, check that the OP
delivery arm is not blocking the chair
movement.
H 52 Patient chair can not be driven upward
because safety switch of side delivery arm
is released.
H 53 Patient chair can not be driven downward
because safety switch of side delivery arm
is released.
H 54 Cup is inserted in cup holder. Remove the cup and make sure the cup fill
tube is above the bowl.
H 55 Disinfectant container attached. To start
waterline cleaning, select instruments and
press Waterline cleaning button.
H 57 Vision action failed. Select patient in
Romexis.
Select patient in Planmeca Romexis before
performing Solanna Vision action.
H 60 WMS: Container almost empty in Domestic
water supply -mode.
Reduce water use for a moment.
H 61 WMS: Container almost empty in Bottled
water -mode.
Reduce water use and fill the container as
soon as possible.
H 62 WMS: Container switch is turned off. Switch on the container switch.
H 63 Cuspidor door is open. Close the door.
H 64 Fill container with disinfectant over upper
sensor.
H 65 Fill container with water over upper sensor.
H 66 Wait 8 hours or switch unit off for the night.
H 67 Instrument flush cancelled. Return
instruments to instrument console.
H 68 Romexis connection lost. Make sure the Ethernet cable is attached
and Romexis is running.
H 69 Unable to create a new user. Maximum
amount of local users exceeded.
31 Help and error messages
260 Planmeca Compact i User's manual