4. Start the short flushing cycle.
NOTE
To be able to run the short instrument flushing cycle from the
Maintenance
window you need an instrument flushing holder. If you do
not have the holder, perform short instrument flushing by running water
through each instrument for 30 seconds / instrument.
Balanced instrument arms: Start the flushing cycle by bending the
instrument arms of those instruments (including syringe) that are to be
flushed to an angle of at least 90° and simultaneously selecting Short
flushing on the control panel.
Hanging-tube instruments: Start the flushing cycle by selecting Short
flushing on the control panel.
The system will first identify the instruments and then each instrument
hose is flushed with air and water for 30 seconds in its turn.
5. Release the instrument arms (if your dental unit is equipped with
balanced instrument arms).
6. Empty the instrument flushing holder.
28 Clean water system
User's manual Planmeca Compact i3 199