7. Repeat the flushing sequence in steps 4 and 5 at least three more times,
or until the water is clean. If the clean-water bottle runs out of water, refill
it.
Empty the instrument flushing holder when needed.
NOTE
If you do not have the instrument flushing holder, run water through each
instrument until the water is clear.
If the water quality of the domestic water is good, you can also use the
domestic water mode when performing the flushing.
8. Balanced instrument arms: Place the instrument hoses back in the
instrument arms, and the instruments back in the instrument console.
Hanging-tube instruments: Place the instruments back in the instrument
holders.
Results
The dental unit is now ready for use.
28 Clean water system