Help messages
HELP
CODE
MEANING ACTION OR EXPLANATION
H 90 This button has no function at the moment,
press Pos Sel, Chair/Headr/Light Sel first.
H 90.1 Measure water chlorine level. Run at least 50 ml of water from the syringe
to the cup. Move the test strip back and
forth in the water for 30 seconds. Compare
the strip colour with the colour scheme.
Select the chlorine level on the control panel
according to the measurement.
< 0.2 (LOW)
0.2 - 0.8 (OK)
> 0.8 (HIGH)
If the chlorine level is continuously too low,
contact service.
H 90.2 ActiveAqua prefilter expired. Switch off the dental unit. Replace the
prefilter. After switching on the dental unit
press the Reset prefilter button to confirm
the prefilter replacement.
To temporarily ignore this notification, select
any instrument or press Cancel.
H 90.3 ActiveAqua water container is empty. Water container fill is slow or the low level
sensor does not sense water.
H 90.4 ActiveAqua water container overflow. Check ActiveAqua level sensors and water
container.
H 90.5 ActiveAqua annual maintenance overdue. Annual ActiveAqua maintenance must be
performed. Contact service.
H 90.6 Measured chlorine level low. Normal use chlorine level increased by
+15%.
H 90.7 Measured chlorine level OK. No change to normal use chlorine level.
H 90.8 Measured chlorine level high. Normal use chlorine level decreased by
-5%.
H 91 Stop (Help) button pressed.
H 92 Help (Stop) button pressed.
H 93 This button has no function at the moment
because there is no motorised light.
H 94 Clock calibration failed. Maximum
calibration range exceeded.
H 95 Patient chair automatic position cannot be
reached because legrest is unlocked or
armrests are blocking movement.
H 96 PlanID tag already in use. PlanID tag assigned to another user.
H 96.1 Different user already logged in. The unit has rejected a login request
because a different user is already logged
in.
H 97 Failed to assign PlanID tag to user. Check Romexis connection.
29 Help and error messages
User's manual Planmeca Compact i3 209

Table of Contents

Other manuals for Planmeca Compact i3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca Compact i3 and is the answer not in the manual?

Planmeca Compact i3 Specifications

General IconGeneral
BrandPlanmeca
ModelCompact i3
CategoryDental equipment
LanguageEnglish

Related product manuals