Remote control commands
R&S
®
NRPxxS(N)
138User Manual 1177.5079.02 ─ 15
The power sensor performs the change of the hostname immediately after the com-
mand is sent. For this purpose, the power sensor restarts its connection to the network,
which can take several seconds. During this time, you cannot address the power sen-
sor. After the restart, you can only address the power sensor using the newly set host-
name.
Note: It is recommended that you do not change the default hostname to avoid prob-
lems with the network connection. However, if you change the hostname, be sure to
use a unique name.
Parameters:
<hostname>
Example:
SYST:COMM:NETW:COMM:HOST
'powersensor-2nd-floor'
Sets powersensor-2nd-floor as new hostname.
SYSTem:COMMunicate:NETWork:IPADdress
<ipaddress>
Effective only:
●
For the R&S NRP LAN power sensors.
●
If SYSTem:COMMunicate:NETWork:IPADdress:MODE is set to STATic.
Sets the IP address of the sensor.
Parameters:
<ipaddress>
Example:
SYST:COMM:NETW:IPAD '192.168.10.29'
Sets 192.168.10.29 as IP address.
Manual operation: See "IP Address" on page 54
SYSTem:COMMunicate:NETWork:IPADdress:GATeway <gateway>
Effective only:
●
For the R&S NRP LAN power sensors.
●
If SYSTem:COMMunicate:NETWork:IPADdress:MODE is set to STATic.
Sets the IP address of the default gateway.
Parameters:
<gateway>
Example:
SYST:COMM:NETW:IPAD:GAT '192.168.10.254'
Sets 192.168.10.254 as IP address of the default gateway.
Manual operation: See "Gateway" on page 54
SYSTem:COMMunicate:NETWork:IPADdress:INFO?
Effective only for the R&S NRP LAN power sensors.