Page #165 background imageLoading...
Page #165 background image
Remote control basics
R&S
®
NRPxxS(N)
165User Manual 1177.5079.02 ─ 15
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
0
Sensor measuring
0
0
0
0
0
0
0
0
0
0
0
0
0
0
+
Figure 11-9: Operation measuring status register
Querying the register:
STATus:OPERation:MEASuring:CONDition?
STATus:OPERation:MEASuring[:SUMMary][:EVENt]?
Table 11-11: Used operation measuring status bits and their meaning
Bit no. Short description Bit is set if
1 Sensor measuring Sensor is measuring.
11.2.5.3 Operation trigger status register
The CONDition register contains information whether a power sensor is waiting for a
trigger event. The EVENt register contains information whether the power sensor has
been waiting for a trigger event since the last query.