Command reference
R&S
®
ZNA
1004User Manual 1178.6462.02 ─ 20
<Mk> Marker number.
Parameters:
<TargetSearchVal> Target search value of marker no. <Mk>.
The value range and reset value depend on the selected target
format (see CALCulate<Chn>:MARKer<Mk>:SEARch:
FORMat on page 1002).
Example:
CALC:MARK ON
Create marker no. 1 and display it in the center of the sweep
range.
:CALC:MARK:TARG -10; FUNC:EXEC TARG
Define a target search value of -10 dB and start the target
search.
CALC:MARK:X?
Query the stimulus value corresponding to the target search
result.
Manual operation: See "Target Value" on page 464
CALCulate<Chn>:MARKer<Mk>:THReshold <Value>
Defines a marker-specific threshold value for (single) peak searches.
Use CALCulate<Chn>:MARKer<Mk>:THReshold:STATe to activate it.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk> Marker number
Parameters:
<Value> Threshold value
The unit is derived from the active trace format and cannot be
modified. Changing the trace format resets the threshold to a
format-specific default value.
Manual operation: See "Threshold Settings" on page 461
CALCulate<Chn>:MARKer<Mk>:THReshold:STATe <Active>
Activates or deactivates the marker-specific threshold for (single) peak searches.
Use CALCulate<Chn>:MARKer<Mk>:THReshold to set the threshold value.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk> Marker number
Parameters:
<Active> Boolean
Manual operation: See "Threshold Settings" on page 461