EasyManua.ls Logo

R&S ZNA67 - Calculate<Chn>:Statistics[:State]

R&S ZNA67
1709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1033User Manual 1178.6462.02 ─ 20
Tip: You can display or hide the "Min/Max/Peak-Peak", "Mean/Std Dev/RMS",
"Phase/El Length" and "Flatness/Gain/Slope" results separately; see example below.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON | OFF - Statistical info field on or off.
*RST: OFF
Example:
*RST; :CALC:STAT:MMPT ON
Reset the instrument, hiding all statistical results. Display the
"Min/Max/Peak-Peak" results.
CALC:STAT:MSTD ON
Display the "Mean/Std Dev" results in addition.
CALC:STAT:RMS ON
Display the "RMS" results in addition.
CALC:STAT:EPD ON
Display the "Phase/El Length" results in addition.
CALC:STAT:SFL ON
Display the "Flatness/Gain/Slope" results in addition.
CALC:STAT:STAT:AREA LEFT, TOP
For a subsequent check at the GUI or a hardcopy, move the info
field to the top left position.
...
CALC:STAT OFF
Hide all results.
Manual operation:
See "Min/Max/Peak-Peak, Mean/Std Dev/RMS" on page 410
CALCulate<Chn>:STATistics[:STATe]:AREA <HorizontalPos>, <VerticalPos>
Moves the statistics info field for the active trace <Chn> to one of nine predefined posi-
tions in the active diagram.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<HorizontalPos> LEFT | MID | RIGHt
Horizontal position
<VerticalPos> TOP | MID | BOTTom
Vertical position
Example: See CALCulate<Chn>:STATistics[:STATe]
Manual operation: See "Min/Max/Peak-Peak, Mean/Std Dev/RMS" on page 410
SCPI command reference

Table of Contents

Other manuals for R&S ZNA67

Related product manuals