Command reference
R&S
®
ZNA
1044User Manual 1178.6462.02 ─ 20
Options: R&S ZNA-K61
Manual operation: See "Compensate Imbalance" on page 649
7.3.1.22 CALCulate:TDVSwr...
CALCulate:TDVSwr[:STATe] <Boolean>
Enables/disables time domain site VSWR measurements.
Parameters:
<Boolean> ON | OFF – TD-VSWR on or off.
*RST: OFF
Manual operation: See "TD-VSWR" on page 405
7.3.1.23
CALCulate:TRANsform...
The CALCulate:TRANsform... commands convert measured data from one repre-
sentation to another and control the transformation into the time domain (see Chap-
ter 4.7.2, "Time domain analysis", on page 220).
CALCulate<Chn>:TRANsform:COMPlex........................................................................ 1044
CALCulate<Chn>:TRANsform:IMPedance:RNORmal...................................................... 1045
CALCulate<Chn>:TRANsform:TIME[:TYPE]................................................................... 1045
CALCulate<Chn>:TRANsform:TIME:CENTer.................................................................. 1046
CALCulate<Chn>:TRANsform:TIME:DCHebyshev.......................................................... 1047
CALCulate<Chn>:TRANsform:TIME:LPASs....................................................................1047
CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam...................................................1047
CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam:CONTinuous............................... 1048
CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam:EXTRapolate.............................. 1049
CALCulate<Chn>:TRANsform:TIME:LPFRequency.........................................................1049
CALCulate<Chn>:TRANsform:TIME:RESolution:EFACtor................................................ 1049
CALCulate<Chn>:TRANsform:TIME:SPAN.....................................................................1049
CALCulate<Chn>:TRANsform:TIME:STARt.................................................................... 1050
CALCulate<Chn>:TRANsform:TIME:STATe.................................................................... 1050
CALCulate<Chn>:TRANsform:TIME:STIMulus................................................................1051
CALCulate<Chn>:TRANsform:TIME:STOP.....................................................................1051
CALCulate<Chn>:TRANsform:TIME:WINDow.................................................................1052
CALCulate<Chn>:TRANsform:TIME:XAXis.....................................................................1052
CALCulate<Chn>:TRANsform:COMPlex <Result>
Converts S-parameters into converted (matched-circuit) Y-parameters or Z-parameters
and vice versa, assuming that port no. i is terminated with Z
0i
so that the three parame-
ter sets are equivalent and the following formulas apply:
SCPI command reference