GUI reference
R&S
®
ZNA
631User Manual 1178.6462.02 ─ 20
In particular, activation is rejected if:
the start frequency is < 31 MHz
the measurement bandwidth is > 100 kHz
the resulting frequency offset (i.e. the minimum frequency offset, rounded to a mul-
tiple of the current frequency step size) would be too high
the port groups are configured with different frequency conversion settings
The generated error code can be retrieved via remote control command
SOURce<Ch>:GROup:SIMultaneous:FOFFset:CONDition?. The configuration
problems listed above are indicated as error codes -8 "invalid offset or frequency spac-
ing" and -6 "no simultaneous mode possible".
Remote command:
SOURce<Ch>:GROup<Grp>:SIMultaneous:FOFFset:CONDition
5.12.3 Pwr Cal Settings tab
Replicates the "Pwr Cal Settings" tab of the "Cal" softtool (see Chapter 5.11.3, "Pwr
Cal Settings tab", on page 572).
5.13 Mode softtool
The mode softtool allows you to prepare the channel for different "non-standard" mea-
surement modes.
Access: Channel – [Mode]
5.13.1 Mode tab
Optimizes the measurement process.
Mode softtool