GUI reference
R&S
®
ZNA
392User Manual 1178.6462.02 ─ 20
Couple All Channels / Decouple All Channels
"Couple All Channels" assigns all traces to the channel of the active trace, deleting
all other (now unused) channels. The analyzer displays a confirmation dialog box
before deleting the unused channels.
"Decouple All Channels" makes sure that each data trace is assigned its own
(independent) channel.
For data traces previously assigned to the same channel, new channels are cre-
ated based on the original channel's settings. Data traces and their associated
memory traces are assigned to the same channel.
Remote command:
n/a
Couple All Scales / Decouple All Scales
"Couple All Scales" couples the scale settings of all traces to the scale settings of
the active trace. The scale settings of the other traces are lost.
"Decouple All Scales" applies independent scale settings to all traces.
If trace A is coupled to trace B, then B's scale settings are copied to A.
Remote command:
n/a
5.5.2 Mem tab
Stores traces to the memory.
Background information
Refer to "Trace types" on page 99.
Coupling of data and memory traces
When a memory trace is generated from a data trace, it is displayed in the same dia-
gram area and inherits all channel and trace settings from the data trace. The memory
trace displayed in the active diagram; its properties are indicated in the trace list: