Command reference
R&S
®
ZNA
887User Manual 1178.6462.02 ─ 20
CALCulate:DATA:DALL? <Format>
Reads the current response values of all data traces of the current recall set. Use
CALCulate:DATA:ALL? to query data traces and memory traces.
Query parameters:
<Format> FDATa | SDATa | MDATa
Output format for the S-parameter data, see
CALCulate<Chn>:DATA.
Return values:
<Data> Response values either in ASCII or block data format, depend-
ing on the current FORMat[:DATA] setting.
Example: Analogous to CALCulate:DATA:DALL?; see
CALCulate<Chn>:DATA.
Usage: Query only
CALCulate:DATA:TRACe <TraceName>, <Format>, <Data>...
The query gets the trace data of an arbitrary (not necessarily the active) trace, refer-
enced by its trace name <TraceName>.
The "set direction" allows to import formatted or unformatted trace data to an existing
trace.
Note
Unformatted trace data (SDATa) can only be imported to memory traces
Formatted trace data (FDATa) can only be imported to "live" traces if the related
channel is in single sweep mode (INITiate<Ch>:CONTinuous OFF).
Before the import, the target trace must be prepared according to the settings used
during export. Any request for new data from the hardware ("Restart Sweep" in sin-
gle sweep mode or switching to continuous sweep mode) discards imported data
and switches back the display to measured data.
Parameters:
<TraceName> String parameter containing the trace name
<Format> FDATa | SDATa | MDATa | NCData | UCData
Data format; see Table 7-2.
<Data> Trace data for FDATa | SDATa import either in ASCII or block
data format, depending on the current FORMat[:DATA] setting.
The column order must match the one used during export.
Example: See CALCulate<Chn>:DATA.
CALCulate<Ch>:DATA:CALL <Format>, <Data>...
The query reads the current response values of all S-parameter data traces at channel
<Ch>'s data access point <DACPoint>.
SCPI command reference