Command reference
R&S
®
ZNA
1174User Manual 1178.6462.02 ─ 20
7.3.9 INSTrument commands
The INSTrument... commands select or query particular resources (SCPI: logical
instruments) of the analyzer.
INSTrument:NSELect................................................................................................... 1174
INSTrument:PORT:COUNt?...........................................................................................1174
INSTrument:TPORt:COUNt?......................................................................................... 1174
INSTrument:NSELect <Channel>
Selects a channel as the active channel.
Parameters:
<Channel> Number of the channel to be activated. The channel must be
created before using CONFigure:CHANnel<Ch>[:STATe] ON.
Range: 1, 2, ...
*RST: 1
Example:
CONF:CHAN2:STAT ON; :INST:NSEL?
Create channel no. 2 and select it as the active channel. The
query returns 2.
INSTrument:PORT:COUNt?
Returns the number of test ports (Port 1, Port 2 ...) of the analyzer.
Example:
INST:PORT:COUN?
Return the number of ports of your analyzer.
Usage: Query only
INSTrument:TPORt:COUNt?
Returns the total number of test ports.
Because the R&S ZNA doesn't support switch matrices, this is always the same as the
number of analyzer ports (see INSTrument:PORT:COUNt?).
Usage:
Query only
7.3.10 MEMory commands
The MEMory... commands control the loaded recall sets of the analyzer.
Storing setups
The MEMory... commands do not affect any stored files. Use the MMEMory... com-
mands to store and load data and to manage files stored on a mass storage device.
SCPI command reference