GUI reference
R&S
®
ZNA
667User Manual 1178.6462.02 ─ 20
Table 5-7: Direct Compensation data
Prompt for Each Port File Type Description
disabled csv One csv trace file per standard,
containing reflection traces for all
selected ports; see Chap-
ter 4.4.2.2, "ASCII (*.csv) files",
on page 157
enabled s1p One 1-port Touchstone file per
standard and port (see Chap-
ter 4.4.2.1, "Touchstone files",
on page 150)
Tip: Remote control provides additional flexibility. You can:
Measure the same ports repeatedly without changing the standards and attribute
the results to different channels.
Calculate the compensation data for different ports, using mixed Open and Short
standards.
Refer to the program example for [SENSe<Ch>:]CORRection:COLLect:
FIXTure[:ACQuire].
Remote command:
[SENSe<Ch>:]CORRection:COLLect:FIXTure:STARt
[SENSe<Ch>:]CORRection:COLLect:FIXTure[:ACQuire]
[SENSe<Ch>:]CORRection:COLLect:FIXTure:SAVE
[SENSe<Ch>:]CORRection:COLLect:FIXTure:EXPort
[SENSe<Ch>:]CORRection:COLLect:FIXTure:IMPort
5.14.3 One Way Loss tab
Defines loss parameters for each physical port.
Use the complementary dock widget to activate or deactivate length/loss compensation
for selected ports (see Chapter 5.14.1.2, "Offset panel", on page 652).
Background information
Refer to the following sections.
Chapter 4.6, "Offset parameters and de-/embedding", on page 201
Chapter 4.6.1.2, "Definition of loss parameters", on page 202
Chapter 4.6.1.4, "Auto Length and Loss", on page 204
Chapter 4.6.1.5, "Fixture Compensation", on page 205
Offset Embed softtool