Welcome to the R&S ZNA
R&S
®
ZNA
19User Manual 1178.6462.02 ─ 20
2 Welcome to the R&S ZNA
This manual is intended to provide you with all information that is necessary for setup,
manual and remote control of the R&S ZNA.
We also invite you to find out what's new in the current revision of the software and
learn how to make best use of our documentation and of the help system.
If you have any questions or comments, please contact your partners at Rohde &
Schwarz and give us your feedback.
2.1 What's new in firmware version 2.41
This manual describes version 2.41 of the R&S ZNA firmware. Compared to the previ-
ous version 2.40, firmware version 2.41 provides the following changes:
New functionality
Leveling datasets can be saved to and loaded from file
See Chapter 5.19.6.4, "Leveling Datasets dialog", on page 833
Eazy de-embedding (R&S ZNA-K210): new "Port Order" setting
See "Port Order" on page 686
New remote control features
New command [SENSe<Ch>:]CORRection:COLLect:AUTO:REPeat to repeat
an automatic calibration
New remote commands DISPlay:LAYout:OVERlay and DISPlay:LAYout:
SPLit for diagram functions Overlay All and Split All, respectively
Improvements
"Power" stimulus axis can be selected without option R&S ZNA-K4.
Solved issues
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:ACQuire
ignored the channel number and used the active channel instead
Touchstone file export: data for 2 frequencies were exported in reverse order
Leveling datasets lost when a power calibration was started
True differential mode (R&S ZNA-K61): with two balanced ports, export of bal-
anced S-parameters raised an error
mm-wave converters: ALC produced wrong levels above a setup-dependent fre-
quency
A [Preset] during an automatic calibration sometimes made the firmware freeze
and crash
µ1 and µ2 in "Stability" parameter selection combo-box erroneously prepended
with "Å"