Command reference
R&S
®
ZNA
1021User Manual 1178.6462.02 ─ 20
h<order>b<recPort>(p<drvPort>)[Relative] of order <order>
Harmonic measurement with:
●
Order <order>
●
Drive port <drvPort>
●
Receive port <recPort>
●
Relative or direct
Cu(P<Src Port>)<_lowercase_ext_dll_name>/<Task
Type>/<Additional Input>
Traces calculated using external DLL <ext_dll_name>.dll.
*) Selecting a parameter Y...<n><m> or Z...<n><m> sets the range of port numbers to
be considered for the Y and Z-parameter measurement to <n>:<m>.
CALCulate<Ch>:PARameter:SDEFine:SENDed <TraceName>, <Result>
Creates a trace and assigns a channel number, a name and a measurement parame-
ter to it. The trace becomes the active trace in the channel but is not displayed.
Similar to CALCulate<Ch>:PARameter:SDEFine, but in presence of balanced ports
this command distinguishes between
●
single-ended S-parameters for logical ports (SSS<Li><Lj>)
●
"raw" single-ended S-parameters referring to physical ports (S<Pi><Pj>)
Suffix:
<Ch>
.
Channel number
Setting parameters:
<TraceName> Trace name
<Result> Measured quantity
Usage: Setting only
Manual operation: See "S-Parameter (selector)" on page 304
CALCulate<Ch>:PARameter:SELect <TraceName>
Selects an existing trace as the active trace of the channel. All trace commands with-
out explicit reference to the trace name act on the active trace (e.g.
CALCulate<Chn>:FORMat). CALCulate<Ch>:PARameter:SELect is also neces-
sary if the active trace of a channel has been deleted.
Suffix:
<Ch>
.
Channel number.
Parameters:
<TraceName> Trace name, e.g. 'Trc4'. See "Rules for trace names" in "Table
Area" on page 390.
SCPI command reference