Command reference
R&S
®
ZNA
1110User Manual 1178.6462.02 ─ 20
Example:
*RST; :CONF:CHAN2:STAT ON; NAME 'New Channel'
Create channel 2 and assign the channel name "New Channel".
CONF:CHAN:CAT?
Query all channels and their names. As a default channel no. 1
is created on *RST, the response is
'1,Ch1,2,New_Channel'.
CONF:CHAN:NAME:ID? 'New Channel'
Query the channel number for the channel named "New Chan-
nel". The response is 2.
Usage: Query only
Manual operation: See "Channel table" on page 601
CONFigure:CHANnel:MEASure:ALL[:STATe] <Boolean>
Enables or disables the sweep in all channels of the active recall set. This command
can be used in combination with CONFigure:CHANnel<Ch>:MEASure[:STATe] to
optimize the measurement speed.
Parameters:
<Boolean> ON | OFF
*RST: ON
Example: See CONFigure:CHANnel<Ch>:MEASure[:STATe]
Manual operation: See "Continuous / Single" on page 511
CONFigure:CHANnel:MEASure:OPTimized <Order>[, <Ch1>, <Ch2>, ...]
This setting tells the firmware to optimize the channel switching times for a particular
channel sweep order.
Note that the optimization can only be applied, if all channels in the current setup are
in single sweep mode (INITiate<Ch>:CONTinuous OFF, and if there is no ongoing
sweep sequence. Any change in the current setup or a wrong sweep sequence lets the
analyzer fall back to unoptimized channel switching.
Setting parameters:
<Order> AUTO | MANual
AUTO
Optimize the switching times for the channel order that is auto-
matically used with INIT:ALL (increasing order of channel
numbers).
MANual
Optimize the switching times for the channel order manually
specified using channel numbers <Ch1>, <Ch2>, ...
SCPI command reference