Command reference
R&S
®
ZNA
1141User Manual 1178.6462.02 ─ 20
Example:
*RST; :DISP:WIND2:STAT ON
Create diagram no. 2.
DISP:WIND2:NAME 'S11 Test Diagram'
Assign a name to the new diagram.
DISP:CAT?
Query all diagrams and their names. As a default diagram no. 1
is created upon *RST, the response is ''1,1,2,S11 Test
Diagram'. The first diagram is not named; its default name is
equal to the diagram number.
CALC:PAR:SDEF 'Win2_Tr1', 'S11'
Create a trace named Win2_Tr1 to measure the input reflection
coefficient S
11
.
DISP:WIND2:TRAC9:FEED 'Win2_Tr1'
Display the generated trace in diagram area no. 2, assigning the
trace number 9 to it.
DISP:WIND2:TRAC:CAT?
Query all traces in diagram area no. 2. The response is
'9,Win2_Tr1'.
Usage:
Query only
Manual operation: See "Title" on page 771
DISPlay[:WINDow<Wnd>]:MAXimize <Boolean>
Maximizes all diagram areas in the active recall set or restores the previous display
configuration.
Suffix:
<Wnd>
.
Number of the diagram area to become the active diagram area.
DISPlay:WINDow<Wnd>:MAXimize acts on all diagrams of
the current recall set, however, the diagram no. <Wnd> is dis-
played on top of the others.
Parameters:
<Boolean> ON | OFF - maximize all diagram areas or restore the previous
display configuration.
*RST: OFF
Example:
*RST; :DISP:WIND2:STAT ON
Create diagram areas no. 1 (with default trace) and 2 (with no
trace).
DISP:WIND2:MAXimize ON
Maximize the diagram areas, placing area no. 2 on top.
Manual operation: See "Maximize Diagram" on page 771
DISPlay[:WINDow<Wnd>]:OVERview[:STATe] <Boolean>
Enables the zoom function with an additional overview window for the diagram no.
<Wnd> or removes the overview window from a diagram.