Command reference
R&S
®
ZNA
1287User Manual 1178.6462.02 ─ 20
Suffix:
<Ch>
.
Channel number of the calibrated channel
Example: See [SENSe<Ch>:]CORRection:COLLect:AUTO:
ASSignment<Asg>:DEFine
Usage: Event
Manual operation: See "Apply" on page 526
[SENSe<Ch>:]CORRection:COLLect:AUTO:TEMPcomp <Boolean>
Available if the calibration unit selected using SYSTem:COMMunicate:RDEVice:
AKAL:ADDRess is an inline calibration system (ICS) only. Applies to all inline calibra-
tion units (ICUs) that are connected to the selected ICS and that are used with factory
characterization data .
If set to ON (1), the R&S ZNA firmware adjusts the factory characterization data
according to the ICU temperatures.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
Manual operation: See "Temperature Compensation" on page 534
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE <CalType>[, <Characterization>[,
<TestPort1>[, <TestPort2>]...]]
Selects and initiates an automatic calibration at arbitrary analyzer and cal unit ports.
This command also selects the calibration type. A progress monitor for the calibration
sweeps is displayed.
Tip: If the test setup contains a high attenuation the analyzer may fail to detect the cal
unit ports connected to each of its ports. In this case use the extended command
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE.
If several calibration units are connected, use SYSTem:COMMunicate:RDEVice:
AKAL:ADDRess to select a unit for the calibration.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<CalType> FNPort | TOSM | FOPort | OPTPort | FRTRans | FTRans |
RTRans | REFL | RSHort | PFNPort | VUOSm | CPUosm
Calibration Type
FNPort
Full n-port (UOSM) calibration
SCPI command reference