Example:
*RST; FREQ:STAR 1E+9; STOP 1.1E+9
Reset the analyzer (activating a frequency sweep) and set the
sweep range between 1 GHz and 1.1 GHz.
SYST:COMM:RDEV:PMET1:DEF 'USB Power Meter',
'NRP-Z55', 'usb', '100045'
Configure an R&S NRP power meter as external power meter
no. 1, assigning the name USB Power Meter and an serial num-
ber 100045.
SENS:FREQ:CONV:ARB:PMET 2, 1, 1E+9, SWE
Convert the receiver frequency to the range between 3 GHz and
3.2 GHz.
Manual operation: See "Frequency Conversion Formula" on page 618
[SENSe<Ch>:]FREQuency<Pt>:CONVersion:AWReceiver[:STATe] <RecFormula>
Defines the frequencies of the "a wave receivers", i.e. the frequencies to which the
receivers of port <Pt> are set while this port is the source port.
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Parameters:
<RecFormula> OFF | ON | SRC
OFF
Only the measurement receiver (b wave receiver) uses the
receiver formula; the reference receiver uses the source for-
mula.
ON
Both receivers use the receiver formula (see [SENSe<Ch>:
]FREQuency<Pt>:CONVersion:ARBitrary.
SRC
Both receivers use the source formula (see SOURce<Ch>:
FREQuency<PhyPt>:CONVersion:ARBitrary:
IFRequency)
*RST:
SRC
Options: R&S ZNA-K4
Manual operation: See "Receiver Freq." on page 616
[SENSe:]FREQuency:CONVersion:DEVice<Port>:NAME <Converter Type>
Sets/queries the type of frequency converter at VNA port <Port>.
Suffix:
<Port>
.
Converter number = VNA port number
SCPI command reference