Command reference
R&S
®
ZNA
1391User Manual 1178.6462.02 ─ 20
Suffix:
<Ch>
.
Channel number
<Seg> Sweep segment number. This suffix is ignored; the setting con-
trols the whole segmented sweep.
Parameters:
<Boolean> ON – use independent power settings, to be defined via
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel].
OFF – reset the power in all sweep segments to the power for
unsegmented sweeps, defined via SOURce<Ch>:
POWer<PhyPt>[:LEVel][:IMMediate][:AMPLitude]. ON
will not restore the previous values.
The parameter is automatically switched to ON when a segment
power is entered using [SENSe<Ch>:]SEGMent<Seg>:
POWer[:LEVel].
*RST: OFF
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using
default settings and thus -10 dBm internal source power.
SEGM:POW -20
Decrease the power to -20 dBm.
SEGM:POW:CONT OFF
Couple the powers in all segments and reset the power in seg-
ment no. 1 to the initial value.
Manual operation: See "Optional Columns" on page 503
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl <MeasDelay>
Defines the delay time for each partial measurement in sweep segment no. <Seg>
("Meas. Delay"). If coupling of the segments is switched on ([SENSe<Ch>:
]SEGMent<Seg>:SWEep:DWELl:CONTrol ON), the delay is valid for all sweep seg-
ments in the current channel.
Suffix:
<Ch>
.
Channel number
<Seg> Sweep segment number
Parameters:
<MeasDelay> Measurement delay before each partial measurement. Changing
the delay leaves the number of points unchanged but has an
impact on the duration of the sweep ([SENSe<Ch>:
]SEGMent<Seg>:SWEep:TIME).
Range: 0 s to 2500 s
*RST: 0 s
Default unit: s
SCPI command reference