Command reference
R&S
®
ZNA
1402User Manual 1178.6462.02 ─ 20
Manual operation:  See "All Partial Meas'ments / First Partial Meas'ment"
on page 490
[SENSe<Ch>:]SWEep:FPOWer <Mode>
Selects the power sweep mode.
Suffix: 
<Ch>
.
Parameters:
<Mode> OFF | ON | FAST
OFF
Regular power sweep
ON
Fast power sweep
FAST
Pulsed power sweep
Manual operation:  See "Fast Power Sweep" on page 498
[SENSe<Ch>:]SWEep:GENeration <mode>
Sets/gets the sweep mode for linear frequency sweeps.
Suffix: 
<Ch>
.
Channel number
Parameters:
<mode> STEPped
The R&S ZNA only supports the stepped mode, i.e, the fre-
quency is changed stepwise and sampling is performed at a
tuned frequency for each measurement point.
*RST:  STEPped
[SENSe<Ch>:]SWEep:POINts <SweepPoint>
Defines the total number of measurement points per sweep ("Number of Points").
Values between 1 and 100,001 can be set.
Suffix: 
<Ch>
.
Channel number 1
Parameters:
<SweepPoint> Number of points per sweep
Range:  see above
*RST:  201
SCPI command reference