[SENSe<Ch>:]SWEep:TIME:AUTO <Boolean>
When enabled, the (minimum) sweep time is calculated internally using the other chan-
nel settings and zero delay ([SENSe<Ch>:]SWEep:DWELl).
Note: The automatically calculated sweep duration is ignored for the sweep types
"Time" and "CW Mode" ([SENSe<Ch>:]SWEep:TYPE).
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON | OFF - turn the automatic calculation of the sweep time on
or off.
OFF is also set if the sweep duration or delay is set explicitly
using [SENSe<Ch>:]SWEep:TIME or [SENSe<Ch>:]SWEep:
DWELl.
*RST: ON
Example:
SWE:TIME 1
Set a total sweep time of 1 s.
SWE:TIME:AUTO?
A query returns the value 1.
Manual operation: See "Sweep Time / Auto" on page 490
[SENSe<Ch>:]SWEep:TIME:MODE <Algorithm>
Defines how the firmware calculates the "hardware settling wait times" in channel
<Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<Algorithm> AUTO | MONO | CONV | SLOW
AUTO
Let the firmware choose the suitable algorithm, based on the
channel's current measurement task
MONO
Manually (and statically) select the algorithm for mono-frequent
measurements
CONV
Manually (and statically) select the algorithm for frequency-con-
verting measurements
SLOW
Manually (and statically) select the algorithm for other "slow"
measurements (e.g. involving power meters)
*RST:
AUTO
Manual operation: See "Timer Settings" on page 651