SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:POWer:STARt <Leveling
Tool Sart Power>
SOURce:POWer<PhyPt>:CORRection:SLEVeling:POWer:STOP <PowValue>
Defines the start and stop power of the leveling grid for physical port <PhyPt>.
After defining the leveling grid (using
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:FREQuency and
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:POWer commands), you
can start the leveling data acquisition using SOURce:POWer<PhyPt>:CORRection:
SLEVeling[:ACQuire].
Suffix:
<PhyPt>
.
Physical port number
Parameters:
<PowValue> Start/stop power
Default unit: dB
Example: See SOURce:POWer<PhyPt>:CORRection:SLEVeling[:
ACQuire]
Options: R&S ZNA-K8
Manual operation: See "Start Power/Stop Power" on page 549
SOURce:POWer<PhyPt>:CORRection:SLEVeling:POWer:STEP <Size>
Defines the (source) power step size of the leveling grid for physical port <PhyPt>.
After defining the leveling grid (using
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:FREQuency and
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:POWer commands), you
can start the leveling data acquisition using SOURce:POWer<PhyPt>:CORRection:
SLEVeling[:ACQuire].
Suffix:
<PhyPt>
.
Physical port number
Parameters:
<Size> Default unit: dB
Example: See SOURce:POWer<PhyPt>:CORRection:SLEVeling[:
ACQuire]
Options: R&S ZNA-K8
Manual operation: See "Power Step Size" on page 549
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:STATe <Active>
Defines whether converter power control is used, if a leveling dataset is available
(SOURce:POWer<PhyPt>:CORRection:SLEVeling:STATe? on page 1437).