For time domain traces, CALCulate<Chn>:MARKer<Mk>:FUNC:DOMAIN:USER
commands did not accept values with units (and returned a misleading error mes-
sage)
For some multichannel configurations, calibration data could not be applied
Markers on traces whose value at the marker position was above or below the visi-
ble range, were not displayed
Frequency converting measurements (R&S ZNA-K4):
Setting the stimulus axis to a converted receiver frequency did not adjust the
frequency range displayed in the channel info
Disabling LO tracking generated an error
Option installation via GUI: Empty alert message if unsuccessful
Calibration setup dialog: "Start Cal" was disabled but no error message was dis-
played, if a combination of cal units and cal kits was configured in a channel (not
supported)
"Calibrate All": If some channels were configured for manual and others for auto-
matic calibration, only the manual calibrations were performed
R&S ZNA firmware PC simulation (R&S ZNXSIM-K2): Non-existing R&S ZNA soft-
ware options displayed in "Info" dialog
Vector mixer measurements (R&S ZNA-K5): Command
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE did not work for cal types
VUOSm and CPUosm
Bad zip compression performance of "Create R&S Support Information" could lead
to GUI freezes and Windows killing the VNA application
Automatic calibration:
A previously defined calibration unit port assignment <Asg> ≥ 2 was not used
in
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:
ACQuire and hence led to bad calibration results
Changing the port assignment to non-default or after auto-detection did not
work and caused an exception
SYSTem:ERRor:DISPlay:STATe OFF did not disable the display of information
popups
"More Ratios" dialog did not work with external generators
Formula-based limit lines were not properly restored from recall sets
With "Same Sweep Setup for All Standards", a combined receiver power calibra-
tion was not performed
Marker format "dB Mag Phase": Incorrect magnitude calculation for certain trace
formats
Frequency converting two-tone group delay measurements (R&S ZNA-K9): Wrong
source frequencies if a source port is not running at channel base frequency
Inapplicable error popup "Sweep time exceeds requested time ..." for time sweeps
with only one point
Calibration kit Keysight 85052C (predefined): Both Reflect standards were defined
as Open instead of Short