GUI reference
R&S
®
ZNA
638User Manual 1178.6462.02 ─ 20
"ZNA and
VSE"
Trace data are processed in the R&S ZNA and I/Q data are pushed
to the R&S VSE.
Note: In the "ZNA and VSE" mode, the R&S ZNA acquires the mea-
surement data for the trace calculation and the I/Q data to be pushed
to the R&S VSE in separate, subsequent sweeps. Make sure that
your DUT can handle the (potentially) doubled pulse sequence with-
out being damaged or changing its characteristics.
Remote command:
CONTrol:VSE:MODE
Trace (b-wave) ← VSE Settings
Selects the b-wave trace to be pushed to the configured R&S VSE instance, if the
selected VSE Mode is different from "ZNA Only".
Remote command:
CONTrol:VSE:TRACe
5.13.2.2 Pulse Modulation dialog
Sets up the analyzer for the modulation, generation and measurement of pulsed sig-
nals.
Mode softtool