GUI reference
R&S
®
ZNA
674User Manual 1178.6462.02 ─ 20
The fixture modeling proceeds in the following steps:
1. Measure one or more test coupons for the related fixture; see Chapter 5.14.5.2,
"Measure Coupon", on page 675
The tools differ in the available coupon types.
2. Measure the DUT with the fixture; see Chapter 5.14.5.3, "Measure DUT + Test Fix-
ture", on page 677
3. Run the selected tool to calculate the Touchstone files modeling the test fixture;
see Chapter 5.14.5.4, "Save Fixture Model", on page 679.
The calculated sNp file(s) can then be used to deembed the DUT at selected
port(s).
All files generated by actions in this dialog are stored in the directory C:\Users\Public
\Documents\Rohde-Schwarz\ZNA\Embedding.
The current implementation assumes symmetrical lead-ins and lead-outs.
5.14.5.1 Common controls on the dialog
The following controls are located at the bottom of each dialog (ISD/SFD/EZD, single-
ended/balanced):
Output Settings
Allows you to specify the directory where the fixture modeling tools store their data.
Default is C:\Users\Public\Documents\Rohde-Schwarz\ZNA\Embedding.
If "Timestamp Measurements" is checked, the names of subsequently generated "Test
Coupon" and "DUT + Test Fixture" files are prefixed with the current date and time.
Remote command:
CALCulate:FMODel:DIRectory
CALCulate:FMODel:DIRectory:DEFault
CALCulate:FMODel:DIRectory:DEFault:CLEar
CALCulate:FMODel:REName
Advanced Settings ...
Opens the "Advanced Settings" dialog of the selected tool. See:
Chapter 5.14.5.6, "ISD Advanced Settings", on page 681
Chapter 5.14.5.7, "SFD Advanced Settings", on page 685
Chapter 5.14.5.8, "EZD Advanced Settings", on page 686
Offset Embed softtool