GUI reference
R&S
®
ZNA
682User Manual 1178.6462.02 ─ 20
Test Coupons > Insertion Loss
Tells the ISD tool about the linearity of the 2xThru test coupon:
"Linear": linear insertion loss
"Non-Linear" (default): non-linear insertion loss
"Resonant": the 2x Thru test coupon is split and used directly for deembedding
This option can be more accurate if the fixture and the 2x Thru have the same
impedance at every location.
Remote command:
CALCulate:FMODel:ISD<Ph_pt>:ATTenuation:BEHavior on page 945
Lead Ins
Defines how the ISD tool interprets the lead-in data.
Scaling for Lead In Atten ← Lead Ins
Scales the test coupon’s attenuation.
Default is 1.
Remote command:
CALCulate:FMODel:ISD<Ph_pt>:SCALe:ATTenuation on page 953