GUI reference
R&S
®
ZNA
758User Manual 1178.6462.02 ─ 20
Remote command:
CALCulate<Chn>:TTIMe:DATA?
5.17.2.10 Skew tab
The "Skew" tab allows you to enable and configure the Skew Measurement.
Access: System – [Applic] > "TDR" > "Skew"
Skew
Enables/disables the skew measurement between the active trace and the Reference
Trace
The skew measurement can only be performed if the following conditions are met for
both the active trace and the reference trace:
the trace format is Real
Time Domain is enabled
the "Low Pass Step" time domain transformation Type is used (which requires the
stimulus grid to be harmonic)
If these conditions are not met, a message is displayed:
Instead of performing the required configuration changes manually, you may also use
the "Auto Setup" button of the corresponding info field to let the firmware do this for
you.