EasyManua.ls Logo

R&S ZNA67 - Page 790

R&S ZNA67
1709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI reference
R&S
®
ZNA
790User Manual 1178.6462.02 ─ 20
Note that for this to work, the port assignments must have a star topology, with a
"Common Port" set using [SENSe<Ch>:]CORRection:COLLect:AUTO:CPORt
on page 1282.
Remote command:
[SENSe:]CORRection:COLLect:AUTO:MCONnect
Same Sweep Setup for All Standards
Selects one of two alternative calibration methods:
If unchecked, then for each measured standard the analyzer individually prepares
the calibration sweep and reduces the number of drive ports to the required mini-
mum.
This method can be faster if the calibration sweeps take longer than the prepara-
tion phase (e.g. due to small bandwidths or many sweep points), or if many ports
are calibrated.
If checked, then each standard involved in an n-port calibration (or in n one-port
calibrations) is measured with n drive ports. The calibration sweep setup is the
same for all standards; no individual preparation phases are required.
This method can be faster if the calibration sweeps are fast compared to the prepa-
ration phases. Due to the similarity of the measurement phases and timing with the
later measurement of the device under test, this calibration method is potentially
more accurate than the previous method. If the sweeps are relatively slow, or if
many ports are calibrated, it causes longer calibration times.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:CSETup on page 1292
Delete Cal Pool / Delete All Cal Kits
Deletes all calibration data and all cal kit data. See Chapter 5.11.4.3, "Calibration Man-
ager dialog", on page 583.
Remote command:
n/a
Search Path for additional Cal Kits and Connector Types
Contains the name and path of a special directory for cal kit files (*.calkit). All cal
kit files in the special directory are loaded automatically as predefined kits (i.e. read-
only kits which cannot be modified) every time the VNA application is started. It is pos-
sible to select the default cal kit directory
C:\Users\Public\Documents\Rohde-Schwarz\ZNA\Calibration\Kits or
any other directory. "None" means that no additional cal kit files are loaded on start-up.
Use the special directory to make sure that you do not have to import kits manually,
even after terminating the VNA application improperly. In this case, previously imported
cal kit files are not stored in the recall set file.
Remote command:
MMEMory:LOAD:CKIT:UDIRectory
Interpolation
Selects the algorithm for the interpolation of user system error corrections.
Setup softtool

Table of Contents

Other manuals for R&S ZNA67

Related product manuals