Command reference
R&S
®
ZNA
954User Manual 1178.6462.02 ─ 20
Manual operation: See "Max Freq to Deembed" on page 684
CALCulate:FMODel:ISD<Ph_pt>:SCALe:FTIMe <FltLeadInScalingTime>
Overrides the lead-in’s flight time in case the through-trace test coupon is a bit too
short or too long.
Sets/gets the atten_scale 1 batch mode parameter of the ISD tool.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<FltLeadInScalingTime>
Manual operation: See "Scaling for Flt Tm" on page 683
CALCulate:FMODel:ISD<Ph_pt>:SMALlfixture <Boolean>
Enables/disables the ISD tool's small fixture mode.
Parameters:
<Boolean>
Manual operation: See "Small Fixture" on page 685
CALCulate:FMODel:ISD<Ph_pt>:SMALlfixture <Boolean>
Enables/disables the ISD tool's small fixture mode.
Parameters:
<Boolean>
Manual operation: See "Small Fixture" on page 685
CALCulate:FMODel:ISD<Ph_pt>:TRACe:COUPling <TraceCoupling>
Tells the ISD tool about coupling among lead-in traces.
Corresponds to the coupling batch mode parameter of the ISD tool.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<TraceCoupling> NONE | WEAK | STRong
NONE: no coupling.
WEAK: coupling will be extracted even if there are 2 ports
enabled on the test coupon
STRong: if the test coupon is a 4-port file, and there are two
ports to be extracted, the ISD tool will optimize odd- and even-
mode insertion losses
Manual operation: See "Trace Coupling" on page 684