EasyManuals Logo

RKC INSTRUMENT PF900 User Manual

RKC INSTRUMENT PF900
170 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #38 background imageLoading...
Page #38 background image
3. WIRING
IMR02L04-E3
3-6
Isolations of the instrument
For isolated device Input/Output blocks, refer to the following:
a
OUT1, OUT2 and OUT3 are isolated when relay contact or triac is specified for OUT1 and OUT2.
b
Not isolated between DO1 and DO2 or DO3 and DO4
c
Digital outputs from DO5 to DO12 are not isolated.
Power supply
Commu-
nication 1
Commu-
nication 2
Digital input
1 to 11
(DI1 to DI11)
Output 1
a
(OUT1)
Output 2
a
(OUT2)
Output 3
a
(OUT3)
Isolated
Digital
output 3
(DO3)
b
Digital
output 4
(DO4)
b
Isolated
Digital
output
5 to12
(DO5 to
DO12)
c
Isolated Isolated
Loader
commu-
nication
Current transformer
input (CT1, CT2)
Feedback
resistance input
(
One of the above)
Measured input
MCU
Non-isolated
Non-isolated
Isolated Isolated Isolated
Isolated
Digital
output 1
(DO1)
b
Digital
output 2
(DO2)
b
Isolated

Table of Contents

Other manuals for RKC INSTRUMENT PF900

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKC INSTRUMENT PF900 and is the answer not in the manual?

RKC INSTRUMENT PF900 Specifications

General IconGeneral
BrandRKC INSTRUMENT
ModelPF900
CategoryController
LanguageEnglish

Related product manuals