MEAS Key - Receiver R&S ESCI
1166.6004.12 4.16 E-2
Selection of the Measurement Function – MEAS Key
The MEAS key opens the menu to select the detectors and set the measurement time for the receiver
and the audio demodulator.
To reduce the amount of data in RFI voltage measurements, a list of subrange maxima (softkey PEAK
SEARCH) can be generated from the scan results and an acceptance line (softkey MARGIN) may be
defined in submenu FINAL MEAS. A final measurement is performed only for frequencies with high RFI
level.
The final measurement may be either automatic (softkey AUTOMATIC FINAL) or in interactive mode
(softkey INTERACTIVE). Automatic control functions for line impedance stabilization networks (softkey
LISN) are available in the side menu. The final measurement is started with the softkey RUN FINAL
MEAS.
The THRESHOLD SCAN function is also used to reduce data. The scan is immediately interrupted
when a limit violation is detected and the final measurement performed within a short time.
MEAS menu:
DETECTOR
MEAS TIME
DEMOD
MEAS
CISPR
AVERAGE
MIN PEAK
MAX PEAK
QUASIPEAK
AVERAGE
QP RBW
UNCOUPLED
AM
DEMOD
ON OFF
FM
FINAL
MEAS
FINAL
PHASES
ESH3-Z5
ESH2-Z5
ENV 4200
OFF
PRESCAN
PHASES
EDIT PEAK
LIST
PEAK
SEARCH
NO OF
PEAKS
RUN FINAL
MEAS
INTER
ACTIVE
AUTOMATIC
FINAL
FINAL
MEAS TIME
PEAKS
SUBRANGES
MARGIN
RUN FINAL
MEAS
THRESHOLD
SCAN
RMS